FLUID COMPONENTS INTERNATIONAL LLC

http://www.fluidcomponents.com

GF Series Flow Meter Models GF90/GF92

36

Doc. No. 06EN003229 Rev. H

To choose a new unit of measure for total flow:

1.

From the Main menu press 3. The 3.0 DISPLAY SETUP menu title will appear.

2.

Press 3, the Totalizer Setup option.

3.

Press 2, Flow Units, to change the units for total flow.

4.

There is the option of standard volumetric units or mass units. See Table 3-4 for a
description of the available choices. Press 1 or 2.

5.

After entering the user password (if it is enabled) the user will be prompted with
standard volumetric or mass units. Enter the desired total flow units.

6.

The user will be prompted to enter the pipe cross-sectional area. The user has the
option of entering the area directly or, for circular pipes, entering the inside diameter.
Enter the method desired.

7.

Enter the desired area units.

8.

Enter the value of the pipe area or the pipe inside diameter. Press any key twice.

9.

When prompted to Save permanently enter Y for yes.

To choose Totalizer Prescaler for total flow (for 4.10 firmware and above):

1.

From the Main menu press 3. The 3.0 DISPLAY SETUP menu title will appear.

2.

Press 3, the Totalizer Setup option.

3.

Press 5, Totalizer Prescaler, to change between prescale values of 0, .001 or 1000.

4.

Press 0 for no prescale, or press 1 for a prescale value of .001, or press 2 for a prescale
value of 1000.

To choose Display Group Sum for total flow (for 4.12 firmware and above):

1.

From the Main menu press 3. The 3.0 DISPLAY SETUP menu title will appear.

2.

Press 3, the Totalizer Setup option.

3.

Press 6, the Display Group Sum.

The totalized flow for the first group will be displayed. Pressing any key will display the totalized
flow for the second group if the second group is enabled. Pressing any key will display the
totalized flow for the third group if the third group is enabled.

The group totals will be displayed regardless if they have been initialized. These values can be
reset by switching to each group and individually resetting each one or doing a reset all from menu
3.3.3.

To choose Aggregate Tot Mode for total flow (for 4.12 firmware and above):

1.

From the Main menu press 3. The 3.0 DISPLAY SETUP menu title will appear.

2.

Press 3, the Totalizer Setup option.

3.

Press 7, Aggregate Tot Mode, to turn on or off the aggregate total mode.

This setting is saved in EEPROM and recalled when the system is reset. When the aggregate mode
is enabled, the following parameters must be the same for all the groups that are linked; totalizer
prescale, totalizer mode (fixed or float) and the totalizer units. If all the parameters do not match, the
totalizer display will show "AGGREGATE TOT ERROR".