To change system parameters in Link Groups mode:

Contact factory to change system parameters in link mode.

NOTE:

The link groups (extended range) and auto-select modes cannot be used simultaneously.
Calibration Functions, Menu 7.0

The calibration functions of the flowmeter are contained in menu level 7.0. These functions are not
needed for general operation and setup of the flowmeter. With the exception of 7.1, Show Delta R, these
functions should be used only after consultation with a factory service representative. Menu level 7.1
displays the RTD resistances and can be compared to calibration resistances. This is useful for
troubleshooting problems and is addressed in Chapter 6, Troubleshooting.

To view RTD resistance values:

1.

Press 7, the Calibration option.

2.

Press 1, the Show Delta R option.

3.

Press 1, for raw dR, 2 for TCdR or 3 for AIdR

Forced Output Setup (EPROM Version 4.20 and later)
Purpose
To force the analog output of the unit.

Password
The user password is required to perform this procedure.

Procedure
Go to the ‘Set Output’ menu by pressing 8, 3, 2. Select the output port you wish to test, either 1 or 2.
Select the percentage of full scale for the output, 0% = 4mA and 100% = 20mA.

To change the forced output, press N for ‘New Value’ or to Exit, press Home.