Doc. No. 06EN003229 Rev. H

61

GF Series Flow Meter Models GF90/GF92

FLUID COMPONENTS INTERNATIONAL LLC

http://www.fluidcomponents.com

Decade Box Verification (example)

The averaging system can be verified using the Delta R calibration sheets and decade boxes to simulate the
delta R signal from the flow elements. The following example begins with the instrument wired for input
power, the Slave flow output is connected the Master Aux 1 input and both instruments are set up using
the same flow units.

The master unit can be switched between the Average Flow on and normal Flow modes to view the results
of the averaging function. This switching is done in the 2.4.2. Menu under the Mode selection.

Master Serial no. 272222

Slave Serial no. 272223

Ref = 1000 ohms Master Display SCFM Master

Display

SCFM

Ref = 1000 ohms Slave Display SCFM

Delta R

Averaging function off Averaging

function

on Delta

R

219.03 0

0

229.46 0

146.16 5

7.775

147.87 10

119.07 25

25.36

129.46 25

95.76 65

57.83

111.35 50

81.38 110

105.2

90.6 100

66.13 190

195.4

68.9 200

54.34 290

295.6

56.54 300

47.66 375

383.1

48.98 390

46.01 400

400.9

45.97 400

Aux’s

Aux4 = 0.000000000
Enabled
Mode = Average Flow

Values for Aux1, Aux2, Enabled and Mode = Average Flow, indicates that the instrument is correctly set up
for averaging flow. Each instrument will have different values for Aux1 and Aux2 depending on the specific
instrument set up.

To return to Main Menu:

Press “Enter”

Press “Home”