Doc. No. 06EN003229 Rev. H

55

GF Series Flow Meter Models GF90/GF92

FLUID COMPONENTS INTERNATIONAL LLC

http://www.fluidcomponents.com

Validation Check Procedure
This procedure systematically guides the user of the VeriCal instrumentation through an initial gathering of
in-situ baseline data. This baseline line data will then be compared to data gathered during similar future
verification processes to determine if the system is operating with-in factory specifications

Recommended Procedure without Pressure Transducer
FCI recommends that this procedure be run during the commissioning process of the instrument to
determine an initial installed baseline calibration and to document any installed offset from the factory
Verical baseline.

Frequency: Every 18 months minimum. The instrument owner has a better understanding for the required
maintenance of their particular process. After the process has been performed several times the customer
should determine the required verification frequency of the process.

This procedure makes the assumption that the instrument has been installed and is completely functional in
the normal operating condition and orientation. The customer should also have access to the factory
VeriCal calibration certificate and the local atmospheric pressure.