Doc. No. 06EN003229 Rev. H

To view the temperature and totalizer units of measure:

1.

From the Main menu press 5. The 5.0 VERIFY menu title will appear.

2.

Press 5 to view the current units for temperature. Press 3 to view the current units for the total
flow. (If the totalizer is disabled the display will only show No Totalizer.) The information will
be displayed to the screen as shown in Figure 24.

TEMPERATURE:

Temp units: °F
Tcal Add = 0.000
Tcal Mult = 1.000

Totalizer Units:

SCF

HIT any key to EXIT

Window 1

Window 2

C00244-1

Figure 24. Temperature, Totalizer Verification Display

Line 2 of screen 1 shows the current units for temperature. Lines 3 and 4 display the offset and
multiplier to the measured temperature. See Advanced Features for a detailed explanation. Screen 2
shows the total flow units.

To view the analog output parameters:

1.

From the Main menu press 5. The 5.0 VERIFY menu title will appear.

2.

Press 6, the Ports option. The two screens in Figure 25 are identical except for the port number.
Line 2 is the current signal output mode. Line 3 shows the full scale flow rate or temperature
value and the appropriate units. Line 4 is the zero value. Press any key for screen 2.

Figure 25. Analog Output Verification

PORT 1

Mode: 4-20 mA
F.S. : 1591 SCFM
ZERO: 0.00 SCFM

PORT 2

Mode: 4-20 mA
F.S. : 150 °F
ZERO: -50 °F

Window 1

Window 2

C00245-1