Specifications – Hach-Lange SI792 C, SI792x C, SI792x C-FF, SI792x C-PA User Manual User Manual
Page 11

Specifications
9
Measurement error
3,4
< 1% measured value + 0.4 µS* c
Temperature input
Temperature sensor
Pt100/PT1000/NTC 100 k
Ω/
NTC 30 k
Ω/NTC 8.55 kΩ (Betatherm)
2-wire connection
Range, Pt100/Pt1000
1
–20.0 to 200.0 °C (–4 to 392 °F)
Range, Pt100/Pt1000
2
–20.0 to 150.0 °C (–4 to 302 °F)
Range, NTC 100 k
Ω
2
–20.0 to 150.0 °C (–4 to 302°F)
Range, NTC 30 k
Ω
–20.0 to 130.0 °C (–4 to 266°F)
Range, NTC 8.55 k
Ω
1
–10.0 to 130.0 °C (14 to 266°F)
Adjustment range
10 K
Resolution
0.1 °C; 0.1 °F
Measurement error
3
,
4
< 0.5 K
(< 1 K for Pt100;<1 K for NTC >100 °C)
Temperature compensation
25 °C reference temperature
1
Applies to si792(x) C and si792x C-FF transmitter only
2
Applies to si792x C-PA transmitter only
3
(± 1 count plus sensor error)
4
IEC 746 Part 1, at nominal operating conditions