Appendix a calibration solutions – Hach-Lange SI792 C, SI792x C, SI792x C-FF, SI792x C-PA User Manual User Manual

Page 89

Advertising
background image

87

Appendix A Calibration solutions

Table 18 Potassium chloride solutions (conductivity in mS/cm)

Temperature

Concentration

1

1

Data source: K. H. Hellwege (Editor), H. Landolt, R. Börnstein: Zahlenwerte und

Funktionen ..., volume 2, part. volume 6

°C

°F

0.01 mol/L

0.1 mol/L

1 mol/L

0

32

0.776

7.15

65.41

5

41

0.896

8.22

74.14

10

50

1.02

9.33

83.19

15

59

1.147

10.48

92.52

16

60.8

1.173

10.72

94.41

17

62.6

1.199

10.95

96.31

18

64.4

1.225

11.19

98.22

19

66.2

1.251

11.43

100.14

20

68

1.278

11.67

102.07

21

69.8

1.305

11.91

104

22

71.6

1.332

12.15

105.94

23

73.4

1.359

12.39

107.89

24

75.2

1.386

12.64

109.84

25

77

1.413

12.88

111.8

26

78.8

1.441

13.13

113.77

27

80.6

1.468

13.37

115.74

28

82.4

1.496

13.62

29

84.2

1.524

13.87

30

86

1.552

14.12

31

87.8

1.581

14.37

32

89.6

1.609

14.62

33

91.4

1.638

14.88

34

93.2

1.667

15.13

35

95

1.696

15.39

36

96.8

15.64

Advertising