4 diagnostic tests, Troubleshooting – Hach-Lange SI792 C, SI792x C, SI792x C-FF, SI792x C-PA User Manual User Manual

Page 78

Advertising
background image

Troubleshooting

76

ERR 33

Sensocheck icon flashes;
Sensoface icon active (see

section 10.2

)

Make sure that the correct
sensor is installed and that it
is operating correctly.
Check sensor connection or
replace cables.
Make sure that there are no
Polarization effects at the
sensor.

Temperature icon flashes;
Sensoface icon active (see

section 10.2

)

Make sure that the
temperature is within the
range of the temperature
compensation table
(temperature compensation,
concentration, salinity).

ERR 98

CONF flashes; configuration
or calibration data is
defective. Memory error in
the program.

Reconfigure and calibrate
the device.

ERR 99

FAIL flashes; EEPROM or
RAM defective

Return to the factory for
repair and calibration.

10.4 Diagnostic tests

Function

Description

View the output
current

From the measuring mode, press

ENTER

. The output

current will show in the main display for five seconds, then
return to measuring mode.

View calibration
data

From the measuring mode, press

CAL

and enter passcode

‘0000’. The cell constant will show in the main display. After
20 seconds, the transmitter will return to the measuring
mode, or press

ENTER

to return immediately to the

measuring mode.

Table 17 Error codes (continued)

Code

Description

Corrective action