3.5.2.1 si792(x) C (4-20 mA/HART) wiring

DANGER
Explosion hazard. Never connect items to the transmitter that are
not specified on the control drawing/ATEX EC-Type certificate.
Refer to

Figure 4

and

Table 1

to connect the power supply to the

si792(x) C transmitter.

Figure 4 si792(x) C (4–20 mA/HART) wiring

1 HART connection (see warnings in

section 3.6 on page 28

)

2 Wiring terminals—see

Table 1

Table 1 Terminal assignments—si792(x) C (4–20 mA/HART)

Terminal No. Assignment

9

Equipotential bonding—ATEX only (Do not use for FM or CSA)

10

4–20 mA output (–)

11

4–20 mA output (+)