3 calibration by comparison or grab sample, Calibration – Hach-Lange SI792 C, SI792x C, SI792x C-FF, SI792x C-PA User Manual User Manual

Page 71

Advertising
background image

Calibration

69

to determine the temperature-corrected conductivity value of the
calibration solution.
Note: If the entry of the calibration solution value takes longer than
6 seconds, the display will alternate between showing the cell
constant and the temperature value.

4. Press

ENTER

to save the value.

5. The display will show the new cell constant value and CELL. Press

ENTER

.

6. The transmitter remains in HoLD mode. Press

ENTER

again.

The transmitter remains in HoLD mode for approximately
20 seconds after calibration and returns to the measure mode.

Note: Repeat the calibration when “Err” is displayed.

Note: The si792x C-PA transmitter cannot be calibrated by this method.

8.1.3

Calibration by comparison or grab sample

The sensor can be calibrated by measuring the process water with a
portable instrument or taking a grab sample and measuring its value in
the lab (one-point calibration). The sensor does not need to be removed
from the process during this calibration. The sample temperature
should correspond to the measured process temperature for accurate
results.
1. Press

CAL

, enter passcode: 1105,

ENTER

. The display will briefly

show CAL PRD and then StorE.

2. Measure the process water and press

ENTER

to proceed. The

current value will be stored, which will be compared to the value
obtained with the lab or portable meter.

The cal mode indicator will flash to indicate that calibration has not
been completed. The transmitter will continue to measure and
display the current sample values.