Operation—4–20 mA/HART

35

Note: During configuration the instrument/transmitter remains on Hold.
Depending on the configuration the loop current is frozen at its last
value or at a preset fixed value.

Press

ENTER

to access a menu item. Use the

ARROW KEYS

to edit

values. Press

ENTER

to save the settings. If a value is outside of the

Select characteristic (Linear / Logarithmic curve)

o1.CHAR

1

LIN (default)

Linear/Logarithmic curve

LOG

Specify the value for the 4 and 20 mA signals
o1. 4mA

000.0 mS

LIN: Enter current start

o1.20mA

000.0 mS

LIN: Enter current end

o1. 4mA

1 mS (default)

LOG: Enter current start

o1.20mA

100 mS (default)

LOG: Enter current end

Set time averaging filter for reducing noise

o1.FtME

0 SEC (default)

Time constant of output filter (refer to

section 5.2.1.2

for more information)

Select a 22 mA signal during errors

o1.FAIL

ON
OFF (default)

Signal behavior during HOLD

o1.HOLD

LAST (default)

During HOLD the last measured value is
maintained at the output

FIX

During HOLD a value (to be entered) is
maintained at the output

Specify the value to output during HOLD periods

o1.FIX

021.0 mA (default)

During HOLD the entered value is
maintained at the output (refer to

section 5.2.1.4

for more information)

1

If the process variables SAL, % (Conc) or USP are selected the following steps

(Lin/Log) are omitted

5.2.1

Output configuration (continued)