Operation—Foundation Fieldbus
43
Select temperature compensation
1
tc.
OFF
Temperature compensation turned off
LIN (default)
Linear temperature compensation
Enter temperature coefficient:
Range: 0–19.99%/K (default: 2.00%/K)
Reference temperature = 25 °C
nLF
Natural waters (to EN 27888)
nACL
Ultrapure water with NaCl traces (0–120 °C)
HCL
Ultrapure water with HCl traces (0–120 °C)
nH3
Ultrapure water with NH
3
traces (0–120 °C)
Select Sensocheck
AL.SnSO
CHECK ON
Continuous Sensocheck evaluation of
sensor function
CHECK OFF (default)
LED in Hold mode
AL.LED
HOLD ON
LED blinks during hold
HOLD OFF (default)
LED off during hold
Enter Fieldbus address (optional)
2
FF.ADR
0026 BUS (default)
Range: 0017–0031
1
Temperature compensation is not available for salinity, concentration or USP.
2
Use only when there is no bus connection. The transmitter will restart and set all
parameters to default values. Individual settings must be entered once more.