Operation—4–20 mA/HART

36

acceptable range, “Err” will be shown and the value will not be
accepted. To exit the menu and return to the measurement mode, press

CONF

and

ENTER

.

Example: Set the output start point for the 4 mA signal to be 10.0 mS
and the output end point to be 100 mS for the 20 mA signal.
1. Press

CONF

, enter passcode:

1200

. The display will show Conf and

then out.1MNU.

2. Press

ENTER

to access the output setup menu. The display will

show

CELL

.

3. Press

ENTER

,

ENTER

,

ENTER

to reach the 4 mA submenu. The

display will show o1.4mA.

4. Use the

UP ARROW

and

RIGHT ARROW

to edit the value to read

10.0 mS value. Press

ENTER

to save the value. The display will

show o1.20mA.

5. Use the

UP ARROW

and

RIGHT ARROW

to edit the value to read

100.0 mS value. Press

ENTER

to save the value. The display will

show o1.FtME.

6. Press

CONF

and

ENTER

to exit the configuration menu.

The transmitter remains in HoLD mode for approximately
20 seconds after calibration and returns to the measurement
mode.

5.2.1.1 Concentration measurement
For the solutions listed in

Table 6

, the transmitter determines the

concentration from the measured conductivity and temperature values
in percent by weight.
The measurement error is made up of the sum of measurement and the
stored accuracy of the concentration curves (see

section B.1 on

page 91

).