1 enter cell constant, 2 calibration with a calibration solution, Calibration – Hach-Lange SI792 C, SI792x C, SI792x C-FF, SI792x C-PA User Manual User Manual

Page 70

Advertising
background image

Calibration

68

8.1.1

Enter cell constant

1. Press

CAL

and enter passcode: 1100,

ENTER

. CAL CELL will be

displayed for 3 seconds.

2. Use the

UP ARROW

and

RIGHT ARROW

to enter the value of the

cell constant of the connected sensor. The conductivity value will be
displayed also.
Note: For Hach or GLI contacting conductivity sensors, the cell
constant is usually printed on a tag with the sensor wiring.

Note: A cell constant change will change the conductivity value.

Note: If the entry of the cell constant takes longer than 6 seconds,
the display will alternate between showing the conductivity and the
temperature value.

3. Press

ENTER

to save the value.

4. The transmitter displays the cell constant value and CELL

.

Press

ENTER

.

5. The transmitter remains in HoLD mode. Press

ENTER

again.

The transmitter remains in HoLD mode for approximately
20 seconds after calibration and returns to the measure mode.

Note: Repeat entry of the cell constant when “Err” is displayed.

8.1.2

Calibration with a calibration solution

Note: Stabilize the temperature during calibration.

1. Press

CAL

, enter passcode: 0110,

ENTER

. CAL SOL will be

displayed for 3 seconds. The transmitter is ready for calibration.

2. Remove and clean the sensor.
3. Immerse the sensor in the calibration solution. Use the

UP ARROW

and

RIGHT ARROW

to enter the value of the calibration solution. If

temperature compensation is OFF, refer to

Appendix A on page 87