Troubleshooting – Hach-Lange SI792 C, SI792x C, SI792x C-FF, SI792x C-PA User Manual User Manual

Page 77

Advertising
background image

Troubleshooting

75

ERR 03

Temperature sensor icon
flashes; open or short circuit;
temperature range exceeded

Test the temperature sensor
wiring.
Make sure that the correct
temperature sensor was
selected in the configuration
menu. For Hach GLI 3400
series sensors use Pt1000.
Measure the resistance of
the temperature sensor in the
sensor to make sure of a
correct reading. For Hach
GLI 3400 series sensors, the
resistance between blue and
white wires should be 1089
to 1106 ohms at 23–27 °C.

ERR 11

Current output icon flashes;
current is below 3.8 mA

Measure the loop power
supply voltage to the
transmitter.

ERR 12

Current output icon flashes;
current is above 20.5 mA

Measure the loop power
supply voltage to the
transmitter.

ERR 13

Current output icon flashes;
reading range value set too
large or too small

Make sure that the 4–20 mA
output range settings cover a
wide enough range and are
correct (see

section 5.2.2

on page 39

).

Table 17 Error codes (continued)

Code

Description

Corrective action

Advertising