Operation—4–20 mA/HART

39

Note: Calibrate a temperature sensor adjustment for a correct
temperature measurement. Use a separate temperature sensor with
fast response for measuring processes with rapid temperature
changes.

Press ENTER to access a menu item. Use the ARROW KEYS to edit
values. Press ENTER to save the settings. If a value is outside of the
acceptable range, “Err” will be shown and the value will not be

5.2.2

Temperature compensation configuration

Select temperature unit

tc.UnIT

°C (default)
°F

Select temperature sensor

tc.rTd

PT100 (default for EU version)
PT1000 (default for standard version)
30 NTC
8.55 NTC

Select temperature compensation

1

tc. LIN

LIN (default)

Linear temperature compensation with entry
of temperature coefficient (00.00–19.99%/
K). Reference temperature = 25 °C.
Enter temperature coefficient
02.00%/K (default) (xx.xx%/K)

nLF

Temperature compensation for naturals
waters to EN 27888

nACL

NaCl: temperature compensation for
ultrapure water with NaCl traces (0–120 °C)

HCL

HCl: temperature compensation for ultrapure
water with HCl traces (0–120 °C)

nH3

NH

3

: temperature compensation for

ultrapure water with NH

3

traces (0–120 °C)

OFF

Temperature compensation turned off

1

If SAL or USP is selected, the following steps are omitted.