2 remove the cartridge from the storage container, Remove the cartridge from the storage container – Hach-Lange NISE sc User Manual

Page 14

Advertising
background image

14

Installation

3.2.2 Remove the cartridge from the storage container

Figure 7

Open the storage container

N O T I C E

Note: This black gasket is not required for installation. It is advisable to keep the black gasket in the
sensor cartridge storage container.

Complete the following movement when you have unpacked the cartridge in order to
moisten the inside of the membranes.

1

Release bayonet coupling

4

Remove black gasket

2

Remove lid

5

Insert black seal in storage container and close the lid.

3

Take out sensor cartridge

The sensor cartridge must not be in contact with air for more than 30 minutes. Make sure that the
electrodes do not dry out.

Advertising
This manual is related to the following products: