3 troubleshooting, 1 troubleshooting during operation, Troubleshooting – Hach-Lange NISE sc User Manual

Page 37: Troubleshooting during operation, 1 troubleshooting, 1 troubleshooting during

Advertising
background image

37

Troubleshooting

6.3 Troubleshooting

6.3.1 Troubleshooting during operation

Symptom

Possible cause

Corrective measures

Incorrect
measurement values

Calibration too old; calibration was not suitable for
the particular application; large change in the waste
water matrix

Perform a suitable calibration.
Refer to

4.6 Calibration/matrix correction,

Page 25

Severely contaminated membranes and/or
reference electrode

Clean the sensor cartridge using a brush
and/or rinse the sensor cartridge with clean
water (without cleaning agents), and wipe the
sensor cartridge carefully with a soft, clean
cloth.

Clean all components (membranes/reference
electrode/temperature sensor).

Install the cleaning unit

Increase the cleaning interval

Sensor membrane damaged

Check the sensor installation/
replace the sensor cartridge

Reference element damaged

NO3 mV RANGE! (Nitrate mV value is out of
measurement range)

1,3

Replace the sensor cartridge

CL mV RANGE! (Chloride value is out of
measurement range)

1,3

REF1 RANGE! (measuring range exceeded on
1st reference value)

REF2 RANGE! (measuring range exceeded on
2nd reference value)

TEMPERATURE (Temperature value is out of
measurement range)

Replace the sensor cartridge/check the
waste water temperature

CARTRIDGE OLD (sensor cartridge more than
1

year

old)

Replace the sensor cartridge

Moisture at the contacts of the sensor cartridge

Dry the contact with a cloth or paper

Check the black gasket for damage and
make sure it is in the correct position.

Screw the 4

socket head screws tight.

Incorrect
measurement values

Moisture inside the measurement probe/faulty
sensor electronics

Check the sensor electronics by using the test
cartridge (

section 7.2, page 39

).

1

Select SENSOR MENU > DIAG/TEST >
SERVICE > TEST CARTRIDGE > Test
cartridge
ready? Press ENTER

2

If all channels are confirmed with OK, the
sensor electronics are operational:
Test cartridge
OK
ENTER

If the test cartridge data is not within this
range and/or if the test cartridge check is not
successful, contact the service department.

Potassium concentrations too high (e.g.: >700

mg/L

in the case of small ammonium concentrations) or
chloride concentrations too high (e.g.: >1000

mg/L
in the case of small nitrate concentrations)

Switch off potassium/chloride compensation
(in the configuration menu - then potentially
enter a fixed value for potassium/chloride)