N O T I C E

18

Installation

3.5 Installation of the probe in the sample flow

There are mounts with separate installation instructions available for installation of the
probe, with or without a cleaning unit, to suit a number of different requirements.

Always observe the following prior to installation:

The probe must be aligned with the bracket, as described in

section 3.5.1, page 18

.

Position the probe at a distance of at least 200

mm (7.87 in.) from the tank wall.

If the probe is attached using a chain mount, make sure that the probe cannot strike
the tank wall.

Immerse the probe at an angle of approx. 45°

±

15°.

Make sure that the probe is fully immersed.

When using the cleaning unit, refer to the supplied instruction sheet.

3.5.1 Position of the probe on the mount

The probe must be attached in a specific position on the mount:

Figure 11

Mount the probe

Handle the sensor cartridge carefully and avoid contact with the membranes when installing the
sensor.

1

Mount the probe on the bracket. 45° adapter and
transition piece should be preassembled.

3

Attaching the aligned probe to the bracket using the
union nut

2

Aligning the probe using the colored reference system
cap. The salt bridge should be oriented upward (12 o'
clock, +/- 5°).

4

Removing the reference system cap