Section 1 technical data, Section 1, Technical data – Hach-Lange NISE sc User Manual

Page 5

Advertising
background image

5

Section 1

Technical data

Subject to change.

General information

AN-ISE sc

AISE sc

NISE sc

Measurement method

Potentiometric measurement using ion-selective electrodes (ISE)

Ammonium and
potassium, nitrate and
chloride, reference
system

Ammonium and
potassium, reference
system

Nitrate and chloride,
reference system

Measurement range

0

to

1000

mg/L

[NH

4

–N]

0

to

1000

mg/L

[K

+

]

0

to

1000

mg/L

[NO

3

–N]

0

to

1000

mg/L

[Cl

]

0

to

1000

mg/L

[NH

4

–N]

0

to

1000

mg/L

[K

+

]

0

to

1000

mg/L

[NO

3

–N]

0

to

1000

mg/L

[Cl

]

Precision

5

% of the measured value + 0.2

mg/L

1

Reproducibility

5

% of the measured value + 0.2

mg/L

1

Response time (90

%)

< 3

minutes

(5

to

50

mg/L)

Measurement interval

Continuous

pH range

pH

5 to pH

9

Calibration methods

Sensor code for sensor cartridge, 1 and 2-point value correction or matrix
correction

Power consumption

1

W

Power supply

Via sc controller

Data transfer

Via sc controller

Ambient data

Typical environment

Used in the biological phase of municipal waste water treatment

Storage temperature

Sensor: –20 to 60

°C (–4 to 140

°F)

Sensor cartridge: 5 to 40

°C (41 to 104

°F)

Operating temperature

Air: –20 to 45

°C (–4 to 113

°F)

Sample temperature

+2 to 40

°C (35 to 104

°F)

Maximum flow velocity

< 4

m/s

Maximum sensor immersion
depth/pressure

Can be immersed to a depth of 0.3 to 3.0

m

(1

to

10

ft);

maximum pressure: 0.3

bar

(4.4

psi).

Maximum Compressed air output
during cleaning unit operation

3.1 bar (45

psi)