3 functional principle, 1 an-ise sc probe, Functional principle – Hach-Lange NISE sc User Manual

Page 9: An-ise sc probe, Order number, General information

Advertising
background image

9

General information

Figure 3

Sensor cartridge

2.3 Functional principle

Ion-selective electrodes have a special membrane to which only a specific type of ion can
adhere. As a result, an ion-specific potential forms on the membrane surface. To measure
a potential difference, a reference system is required that will not be affected by the
sample to be measured.

The CARTRICAL

TM

technology reduces cross-sensitivity by calibrating not only the

individual electrodes but also the measuring electrode against the compensation
electrode and the reference; this is carried out at the factory. The reference system is
designed using pH-differential technology and is therefore particularly stable in terms of
drift and contamination.

2.3.1 AN-ISE sc probe

The AN-ISE sc probe uses the ion-selective electrode technology to measure ammonium
ions (NH

4

+

) and nitrate ions (NO

3

) in a waste water sample.

Known interfering factors due to potassium (when measuring ammonium), chloride (when
measuring nitrate) and temperature are compensated by suitable built-in electrodes.

1

Ammonium electrode

1,2

1

Active with AN-ISE sc

2

Active with AISE sc

5

Reference system

2

Nitrate electrode

1,3

3

Active with NISE sc

6

Temperature sensor

3

Potassium electrode

1,2

7

Marker hole for assembling the probe

4

Chloride electrode

1,3