Operation – Hach-Lange NISE sc User Manual

Page 22

Advertising
background image

22

Operation

SENSOR MENU

AN-ISE sc or AISE sc or NISE sc

CALIBRATE

MATRIX CORR

Matrix correction options.
The most recently used menu is displayed.
The currently active corrections are shown in Information.

NONE

No MATRIX CORR. is activated

MATRIX 1

1-point matrix correction

NH4 + NO3

1

1 point matrix correction for ammonium and nitrate

NH4

1,2

1-point matrix correction for ammonium

NO3

1,3

1 point matrix correction for nitrate

NH4 + K

1,2

1 point matrix correction for ammonium and potassium

NO3 + CL

1,3

1 point matrix correction for nitrate and chloride

NH4+K NO3+CL

1

1 point matrix correction for ammonium, potassium, nitrate and
chloride

TAKE SAMPLE IMMEDIATELY AND
ANALYSE IN LABORATORY

Information window: When this window appears, the sample must
be taken immediately and then analyzed in the laboratory.

VALUE CORR. 1

Perform 1 point value correction.

The most recently used menu is displayed.
The currently active corrections are shown in Information.

NH4–N

1

Select parameter for the 1 point value correction

NO3–N

1

VALUE POINT

Enter the values for the 1 point value correction

Note: The following example shows the input when using a
AN-ISE sc probe for ammonium. When using the AISE sc probe,
the input is the same. When using the NISE sc, it is only possible to
input the nitrate and chloride values.

AN-ISE SC NH4–N

Enter the displayed ammonium value

AN-ISE SC K

Enter the displayed potassium value

LAB NH4–N

Enter the laboratory ammonium value

ENTRY COMPLETE

Confirm the values entered

CORR-RESULT

Display the correction results

VALUE CORR. 2

Perform 2-point value correction

NH4–N

1

Select parameter for the 2 point value correction

NO3–N

1

VALUE POINT 1

Enter the values for the 2 point value correction (first point)

Note: The following example shows the input when using a
AN-ISE sc probe for ammonium. When using the AISE sc probe,
the input is the same. When using the NISE sc, it is only possible to
input the nitrate and chloride values.

AN-ISE SC NH4–N

Enter the displayed ammonium value

AN-ISE SC K

Enter the displayed potassium value

LAB NH4–N

Enter the laboratory ammonium value

ENTRY COMPLETE

Confirm the values entered