Troubleshooting

6.2 Warnings

I

n the event of a sensor warning, all menus, relays and outputs continue to function as

normal but a warning symbol lights up.

Warnings may be used to activate a relay; users can set warning levels to define the
severity. Warnings are defined in

Table 3

.

Table 3

Warnings

Displayed warnings

Cause

Resolution

RFID DATA

Cartridge faulty, read process failed

Replace cartridge, check probe
with test cartridge

NH4 mV RANGE!

1,2

Ammonium mV value is close to measuring range limit

Refer to

6.3.1 Troubleshooting

during operation, Page 37

.

K+ mV RANGE!

1,2

Potassium mV value is close to measuring range limit

NO3 mV RANGE!

1,3

Nitrate mV value is close to measurement range limit

Cl

mV RANGE!

1,3

Chloride mV value is close to measurement range limit

REF1 mV RANGE!

1st reference value is close to limit

REF2 mV RANGE!

2nd reference value is close to limit

TEMPERATURE

Temperature is close to limit

CARTRIDGE OLD

Sensor cartridge more than 1 year old

Replace the sensor cartridge

NH4-N CONC HIGH

1,2

Ammonium concentration value exceeds measuring range

See

6.3.1 Troubleshooting during

operation, Page 37

.

NH4-N CONC LOW

1,2

Ammonium concentration value is below measuring range

NO3-N CONC HIGH

1,3

Nitrate concentration value exceeds measuring range

NO3-N CONC LOW

1,3

Nitrate concentration value is below measurement range

K+ KONZ HOCH

1,2

Potassium concentration value exceeds measuring range

K+ CONC LOW

1,2

Potassium concentration value is below measuring range

CL CONC HIGH

1,3

Chloride concentration value exceeds measuring range

CL CONC LOW

1,3

Chloride concentration value is below measurement range

AMMONIUM

1,2

Refer to

6.3.2 Troubleshooting

during calibration, Page 38

.

OFFSET

Ammonium offset exceeds measurement range

SLOPE

Ammonium slope exceeds measurement range

POTASSIUM

1,2

OFFSET

Potassium offset exceeds measurement range

SLOPE

Potassium slope is out of measuring range

NITRATE

1,3

OFFSET

Nitrate offset is out of measurement range

SLOPE

Nitrate slope is out of measurement range

CHLORIDE

1,3

OFFSET

Chloride offset is out of measurement range

SLOPE

Chloride slope is out of measurement range

1

Applies to AN-ISE sc

2

Applies to AISE sc

3

Applies to NISE sc