Operation

4.6.4 Performing the matrix correction

Note: Take laboratory value measurements or reference values promptly or, alternatively, take these
from the stabilized sample. This will prevent changes in sample concentration, as time is a factor in
comparative tests.

Refer to

7.3 Validation accessories, Page 39

for recommended laboratory measurement

tests.

4.6.4.1 MATRIX 1 correction (1 point matrix correction)

Proceed as follows to perform MATRIX 1:

1.

Select

SENSOR MENU > AN-ISE SC

or

AISE SC

or

NISE SC > CALIBRATE > MATRIX

CORR.

.

2.

Select

MATRIX 1

from the selection window and press

ENTER

.

3.

Select the parameters you wish to correct and confirm by pressing

ENTER

.

Selection options for AN-ISE sc:
NH

4

+ NO

3

; NH

4

; NO

3

; NH

4

+ K; NO

3

+ Cl; NH

4

+ K NO

3

+ Cl

Selection options for AISE sc:
NH

4

; NH

4

+ K

Selection options for NISE sc:
NO

3

; NO

3

+ Cl

The sensor saves the current values of the selected parameters at this point.

4.

Take a water sample immediately from the closest point possible to the sensor. Filter
the sample as quickly as possible and carry out a prompt laboratory analysis of the
selected parameters, as the measurement value can change quickly.

When the laboratory value has been determined, proceed as follows:

5.

Select

SENSOR MENU > AN-ISE SC

or

AISE SC

or

NISE SC > CALIBRATE > ENTER

LABVALUE

.

6.

The laboratory values for the parameters can only be entered if the MATRIX1
correction has been selected beforehand. Once the laboratory values have been
entered, select

ENTRY COMPLETE

to confirm.

When the entered laboratory value is confirmed, the matrix correction is activated.

7.

Once the correction is activated, the result

CORR-RESULT

is shown.

Note: This process must always be carried out in full to make sure the matrix correction is
completed successfully.
If a correction does not produce a successful result, calculations are made with the previous
correction.

CALIBRATE

MATRIX CORR

FURTHER CORR.

INFORMATION

MATRIX 1

TAKE SAMPLE

IMMEDIATELY

AND ANALYSE

IN LABORATORY

CALIBRATE

MATRIX CORR

FURTHER CORR.

LABORW: EING.

INFORMATION