Operation

4.6 Calibration/matrix correction

The four electrodes with the reference system of the compact sensor cartridge were
calibrated with one another at the factory using special standard solutions (CARTICAL

TM

).

However, the membranes on the ion-selective electrodes are not 100%

selective due to

other substances that may affect the measurement. Perform a matrix correction (refer to

4.6.4, page 28

) to compensate for other ions present on the ISE electrodes.

Potassium has the largest interference effect on the ammonium membrane, while chloride
has the largest effect on the nitrate membrane. The AN-ISE

sc probe compensates for this

problem with the aid of a built-in potassium/chloride electrode.

When using the AISE sc probe, only the ammonium membrane and the integrated
potassium electrode are active.

SENSOR MENU

SIGNALS

Signals and measurement results of the individual measurement
channels

AMMONIUM

1,2

Display the signals and measurement results for ammonium

NITRATE

1,3

Shows the signals and measurement results for nitrate

POTASSIUM

1,2

Display the signals and measurement results for potassium

CHLORIDE

1,3

Display the signals and measurement results for chloride

REF. ELECTRODE

Displays the signals and measurement results for the reference
system

MV RAW

Display the signals and measurement results for MV RAW

IMPED STATUS

Display the signals and measurement results for impedance

TEMP

Display the signals and measurement results for temperature

HUMIDITY

Display the signals and measurement results for humidity

RFID

Display the signals and measurement results for RFID

CAL DAYS

Shows the age of the last matrix correction

AMMONIUM

1,2

Display the age of the last matrix correction for ammonium

NITRATE

1,3

Display the age of the last matrix correction for nitrate

SERVICE

TEST CARTRIDGE

Perform a sensor check with the test cartridge

TEST CARTRIDGE READY? PRESS ENTER

TEST CARTRIDGE

Display whether the individual sensor channels are OK or not

DIAG/TEST

Displays whether or not DIAG/TEST is OK

GNDROD

Displays whether or not GNDROD is OK

REF

Displays whether or not REF channel is OK

NO3

1,3

Displays whether or not NO3 channel is OK

NH4

1,2

Displays whether or not NH4 channel is OK

ORP

Displays whether or not ORP channel is OK

CL

1,3

Shows whether or not Cl channel is OK

K+

1,2

Displays whether or not K channel is OK

TEMP

Displays whether or not temperature channel is OK

CHANGE CARTR.

Follow the menu process

CLEANING

Follow the menu process

1

Applies to AN-ISE sc

2

Applies to AISE sc

3

Applies to NISE sc