2 troubleshooting during calibration, Troubleshooting during calibration, 2 troubleshooting – Hach-Lange NISE sc User Manual
Page 38: Troubleshooting, 1 troubleshooting during operation (continued)

38
Troubleshooting
1 applies to AN-ISE sc
3 applies to NISE sc
Unstable
measurement values
Air bubbles, depth of immersion
Check the sensor installation
Check the cleaning unit configuration
Moisture at the contacts of the sensor cartridge
Dry the contacts with a cloth or paper.
Check the black gasket for damage and
make sure it is in the correct position.
Screw the 4
socket head screws tight
Sensor membrane damaged
Check the sensor installation/
replace the sensor cartridge
Reference element damaged
6.3.1 Troubleshooting during operation (Continued)
Symptom
Possible cause
Corrective measures
6.3.2 Troubleshooting during calibration
Symptom
Possible cause
Corrective measures
SENSOR CODE
Sensor code entered incorrectly
Using the certificate, check whether the
sensor code was entered correctly.
AMMONIUM
1, 2
OFFSET
Error during the last ammonium correction, sensor
cartridge too old, contaminated, faulty
Repeat the correction.
Use the previous correction.
Clean or replace the sensor cartridge.
SLOPE
POTASSIUM
1,2
OFFSET
Error during the last potassium correction, sensor
cartridge too old, contaminated, faulty
Repeat the correction.
Use the previous correction.
Clean or replace the sensor cartridge.
SLOPE
NITRATE
1,3
OFFSET
Error during the last nitrate correction, sensor
cartridge too old, contaminated, faulty
Repeat the correction.
Use the previous correction.
Clean or replace the sensor cartridge.
SLOPE
CHLORIDE
1,3
OFFSET
Error during the last chloride correction, sensor
cartridge too old, contaminated, faulty
Repeat the correction.
Use the previous correction.
Clean or replace the sensor cartridge.
SLOPE
1
Applies to AN-ISE sc
2
Applies to AISE sc
3
Applies to NISE sc