23

Operation

SENSOR MENU

VALUE POINT 2

Enter the values for the 2 point value correction (second point).

Note: The following example shows the input when using a
AN-ISE sc probe for ammonium. When using the AISE sc probe,
the input is the same. When using the NISE sc, it is only possible to
input the nitrate and chloride values.

AN-ISE SC NH4–N

Enter the displayed ammonium value

AN-ISE SC K

Enter the displayed potassium value

LAB NH4–N

Enter the laboratory ammonium value

ENTRY COMPLETE

Confirm the values entered

CORR-RESULT

Display the correction results

FURTHER CORR.

Other matrix correction options

None

No

FURTHER CORR.

is activated

MATRIX 2

A 2-point matrix correction can be performed here

NH4

1

Parameter selection for the MATRIX2 correction.

NO3

1

MEAS CONC 1

Saves the currently measured measurement for the first point

DATE

Displays the date of the current correction of the first point

CONC. LABVALUE 1

Entry and display of reference value for the first point

MEAS CONC 2

Saves the currently measured measurement for the second point

DATE

Displays the date of the current correction of the second point

CONC. LABVALUE 2

Entry and display of the reference value for the second point

HIST. CORR.

Selection of one of the last corrections performed

SENSOR CODE

The sensor code can be activated or entered here

ACTIVATION

Activates the sensor code for the individual channels

NH4 + K

1

Activate the sensor code for ammonium and potassium

NO3 + CL

1

Activate the sensor code for nitrate and chloride

NH4+K NO3+CL

1

Activate the sensor code for ammonium, potassium, nitrate and
chloride

FACTORY CALIBRATION

Activates factory calibration

INPUT

Entry of the sensor code

ENTER CORR.

The laboratory values of the last matrix correction can be changed

ENTER LABVALUE (displayed when MATRIX 1 or
MATRIX 2 is performed)

Enter the laboratory values if MATRIX 1 or MATRIX 2 has been
selected

AMMONIUM

1,2

Entry of ammonium laboratory value

NITRATE

1,3

Enter the laboratory value for nitrate

POTASSIUM

1,2

Enter the laboratory value for potassium

CHLORIDE

1,3

Enter the laboratory value for chloride

ENTRY COMPLETE

Confirm the values entered

CORR-RESULT

Display the correction results

NH4–N

1,2

Displays whether or not the ammonium correction was successful

NO3–N

1,3

Shows whether or not the nitrate correction was successful

K+

1,2

Displays whether or not the potassium correction was successful

CL

1,3

Displays whether or not the chloride correction was successful