Ensure that the correct control lot number and target values were
selected and entered correctly.

•

Verify that the system is functioning properly by running the
system controls.

•

If the system controls pass, contact the kit manufacturer.

Sample High/Low

This message occurs if the sample results do not fall within the
highest and lowest standards.

•

Ensure that you selected the correct template.

•

Ensure that the correct standard lot number and target values
were selected and entered correctly.

•

Verify that the system is functioning properly by running the
system controls.

•

If the sample is noted as High, dilute the sample following the
kit manufacturer’s suggestions.

•

If the system controls pass, contact the kit manufacturer.

Failed Standard in Batch

This message occurs when one or more of the assay standards are not
within the expected range.

•

Ensure that you selected the correct template.

•

Ensure that the correct assay standard lot number and target
values were selected and entered correctly.

•

Verify that the system is functioning properly by running system
controls.

•

If the system controls pass, contact the kit manufacturer.

Sample Timed Out

•

Ensure that the samples were properly mixed before they were
dispensed into the wells.

•

Ensure that you loaded the sample to the correct well on the
plate.

•

Check if the sample volume added to the well is correct.

•

Remove and sonicate or flush the XYP sampling probe to
remove possible clogs.

•

Check if the XYP sampling probe is properly aligned.

•

Complete the following analyzer functions from the Maintenance
tab: Drain, Backflush, and Alcohol Flush.

•

Run the system controls.

•

Check to see if the delta cal temperature < 3.0 degrees.
Recalibrate if greater than 3.0 degrees and retry.

•

If the system controls pass, contact the kit manufacturer.