Batches, Templates – Luminex 100 IS User Manual, Version 2.1 User Manual
Page 98
100 IS S Users Manual Version 2.1
x
MAP Technology
7 - 34
PN 89-00002-00-070 Rev. A
click that row. An error dialog box appears with details about that 
error.
Batches
A batch is a group of samples that will be processed using a single 
template. Batches consist of templates, lots, and samples for 
acquisition and can span more than one plate.
Templates are provided by the assay kit manufacturer via CD ROM 
or floppy disk. When creating a new batch, you must use a template. 
Information that you enter does not override or change the template.
The kit manufacturer includes reagent assays in your assay kit. The 
system requires you to provide information about the reagent assays 
you want to use in the batch.
A batch can include more than one plate. If you run out of wells on 
the microtiter plate, another plate appears for additional samples. The 
new plate appears to the immediate right of the existing plate image 
on the screen with a dark line between the adjacent columns of the 
two plates.
Templates
When creating batches or multi-batches, you must select a template 
to associate with the batch. 
Templates are pre-defined sequences of commands that an kit 
manufacturer has determined must be performed during each batch 
acquisition.
Import a Template
Your assay kit provides a template on a diskette specific to the kit. 
You need to import new templates to the system only once. 
Templates can include standards or controls, maintenance 
commands, and patient acquisition commands.
To import a template from diskette:
1. Insert the kit’s diskette into the disk drive.
2. Click the Import Template button on the toolbar. This dialog
box appears.