Luminex 100 analyzer problems, Sample empty detected, Unknown formula failure – Luminex 100 IS User Manual, Version 2.1 User Manual

Page 159: Power on indicator light is not on, Pc is not powered on, Sheath fluid is leaking onto the plate holder

Advertising
background image

x

MAP Technology

Troubleshooting

PN 89-00002-00-070 Rev. A

9 - 9

Sample Empty Detected

This error message occurs when the system completely acquires the
entire sample volume.

•

See if the samples were properly mixed before they were
dispensed into the wells.

•

See if you loaded the sample to the correct well on the plate.

•

Check if the sample volume added to the well is correct.

•

Remove and sonicate or flush the XYP instrument Sample Probe
to remove possible clogs.

•

Check if the XYP instrument Sample Probe is properly aligned.

•

Complete the following analyzer functions from the Maintenance
tab: Drain, Backflush, and Alcohol Flush.

•

Run the system controls.

•

Check to see if the delta cal temperature < 3.0 degrees.
Recalibrate if greater than 3.0 degrees and retry.

•

If the system controls pass, contact the kit manufacturer.

Unknown Formula Failure

•

Ensure that you selected the correct template.

•

Verify that the system is functioning properly by running system
controls.

•

If the system controls pass, contact the kit manufacturer.

Luminex 100

Analyzer Problems

Power On Indicator Light
Is Not On

•

Ensure that the Luminex 100 analyzer is powered on.

•

Ensure that the system is plugged into an outlet with current.

•

If using a surge protector, see if the power strip is On.

•

Remove and reconnect the cable.

•

Check the fuses (see “Fuses” on page 8-10).

PC Is Not Powered On

•

Ensure that the PC voltage is set correctly by checking the back
of the PC. The options are 115 Volts and 230 Volts.

•

Ensure that the system is plugged into an outlet with current.

•

If using a surge protector, see if the power strip is On.

•

Remove the cable and reconnect.

•

Power on the PC.

Sheath Fluid Is Leaking
Onto the Plate Holder

•

Remove and sonicate or flush the XYP instrument Sample
Probe. Load a new plate onto the plate holder.