Luminex 100 IS User Manual, Version 2.1 User Manual
Page 55
x
MAP Technology
The System
PN 89-00002-00-070 Rev. A
5 - 13
Air, Waste, and
Sheath Fluid
Connectors
Located on the side of the instrument, these connectors couple 
directly to the sheath and waste fluid containers. The air connector is 
green, the sheath fluid connector is blue, and the waste fluid 
connector is orange.
1. Sheath Connector (Blue)
2. Air Connector (green)
3. Waste Connector (orange)
Figure 32. Air, Waste, and Sheath Fluid Connectors
If your system uses the Luminex Sheath Delivery system, see 
“Luminex Sheath Delivery System” in the appendix for more 
information about the Sheath Delivery system and its installation.
Sheath Fluid
Container (not
shown)
For proper operation, this container must be placed at the same 
level as the base of the XYP instrument. The sheath fluid container 
should not be placed on top of the Luminex 100 analyzer. The sheath 
fluid level should be below the air inlet connection and above the 
sheath outlet connection.
Waste Fluid
Container (not
shown)
This receives waste from the system. Waste levels must be 
manually monitored—do not allow the waste container to 
overflow! Empty the waste container each time the sheath fluid 
container is filled. The waste container should not be placed on top 
of the instrument.
1
3
2
Note:
Sheath fluid levels
must be monitored 
manually. Check the 
sheath fluid level before 
starting a run or 
procedure.
Warning:
If biological samples have been tested with the system,
use your standard laboratory safety practices.