Luminex 100 IS User Manual Version 2.1

x

MAP Technology

4 - 2

PN 89-00002-00-070 Rev. A

instrument. Do not plug the XYP instrument into the power
outlet.

5. Place the Luminex 100 analyzer onto the XYP instrument.

6. Attach the power cords to the input module of the Luminex 100

analyzer and attach the USB cable to port 1 on the Luminex 100
analyzer. Do not plug into the power outlet. Make sure that the
power switches to the Luminex 100 analyzer and to the XYP
instrument are off. See connections photos on page 4-3 or the
diagram on page 4-4.

7. Install the alignment guide (translucent plastic tube) by screwing

it into the XYP instrument where shown. Adjust the position of
the Luminex 100 analyzer with the XYP instrument until the
alignment guide threads screw in completely.

Figure 14. Alignment Guide

8. Set the voltage selection switch on the PC’s power supply to

match the ac power available at your location to prevent damage
to your PC.

Set it to 115 volts (V) in most of North and South America and
some Far Eastern countries such as Japan, South Korea, and
Taiwan.

Set it to 230 volts (V) in most of Europe, the Middle East, and
the Far East.

Alignment
guide

Note:

The ac power

requirements may vary from
location to location and region to
region.