Xyp instrument, Pc specifications, Electronics – Luminex 100 IS User Manual, Version 2.1 User Manual

Page 48: General

Advertising
background image

Luminex 100 IS User Manual Version 2.1

x

MAP Technology

5 - 6

PN 89-00002-00-070 Rev. A

Electronics

•

Reporter channel detection: A/D resolution 14 bits

•

Communications interface: USB

•

XYP instrument, communications interface: RS 232

XYP Instrument

General

•

Ambient temperature: 15°C to 30°C (59°F to 86°F)

•

Humidity: 20% to 80%, noncondensing

•

Altitude: operation up to 2400 m (7874 ft) above mean sea level

•

Physical dimensions: 44 cm (17.25 inches) W x 60 cm (23.5
inches) D x 8 cm (3 inches) H

•

Weight: 15 kg (33 lbs.)

•

UL installation category: UL Installation Category II, as defined
in Annex J of UL 61010A-1

•

Pollution degree: UL Pollution Degree 2, as defined in Section
3.7.3.2 of UL 61010A-1

•

Heater operating range: 35°C to 55°C (95°F to 131°F) with
tolerance 0°C to +2°C

•

XY Platform, input voltage range: 100 - 240 V~, 1.8 Amps,
47 - 63 Hz

•

AC inlet fuse: 3 Amp, 250 Vac, fast acting

PC Specifications

•

Computer: a Pentium

®

III or equivalent class PC, minimum

speed 933 MHz

•

Main memory: minimum of 256 MB

•

Hard disk drive: 10 GB minimum storage capacity

•

Communications, serial ports: Minimum of 2 RS-232 compatible
serial ports

•

Communications, parallel pointing device and serial ports:
Minimum of 1 parallel, 1 PS/2 compatible pointing device, and 2
RS-232 compatible serial ports

•

Communications, USB port: Minimum of one USB Version 1.1
compatible high speed port

•

Read, write removable media: CDR/W-type drive, standard 1.44
MB 3.5-inch disk drive

•

Operating system support: MS Windows

®

•

CE marked and UL listed

•

Two-button mouse or equivalent

•

Monitor: A minimum of 17-inch diagonal monitor

•

Keyboard: a 104-key keyboard or equivalent

Advertising