Needle / electrode resistance testing, Barrel removal, Ransburg – Ransburg Vector Solo 79901 Waterborne RS90-WB User Manual

Page 34

Advertising
background image

30

Vector Solo Waterborne Applicators - Maintenance

Ransburg

NEEDLE / ELECTRODE

RESISTANCE TESTING

The electrical resistance of the needle/electrode

should be tested periodically (typically on a weekly

basis) or any time it is removed from the applicator.

To Test

1. Install the needle/electrode onto the front end

of an available needle shaft. Be sure that the

needle/electrode is completely seated for proper

contact between the metal shaft and the threaded

insert of the needle/electrode.

2. Using a VOM meter that will read 15 megohms

accurately, connect the first meter lead to the metal

needle shaft and the second lead to the needle/

electrode wire. The needle/electrode resistance

should be 14.5 to 19 megohms (nominal 15 meg-

ohms at 9 volts or 12 to 17 megohms at 1000

volts). Needle/electrodes outside these ranges

must be replaced.

Figure 28: Testing Resistive Electrode

BARREL REMOVAL

1. Remove air cap and fluid nozzle.

2. Remove two (2) screws with a 3mm Allen wrench

that hold the rear cover. Remove rear cover and

disconnect the connector.

Figure 29: Rear Cover Removal

3. Remove the rear fluid adjustment bushing and

air valve and fluid needle return springs.

Figure 30: Fluid Adjust Bushing Removal

AH-08-03.6

Advertising