Rma-303 indirect charge - maintenance, Fluid tube removal/replacement – Ransburg RMA Indirect A12869 User Manual

Page 70

Advertising
background image

RMA-303 Indirect Charge - Maintenance

65

LN-9268-11.4

Figure 30: Reinstall Fluid Tip

Figure 31: Fluid Tip/Tube Gap

!

W A R N I N G

Figure 32: Fluid Tip Removal

Replacement
Ensure the tip openings are fully open and
clean. Apply an O-ring lubricant to the O-ring
to help hold it in place on the fluid tip. Insert
the O-ring into the undercut groove on the tip.
Place the tip on the tool and tighten in a coun-
ter-clockwise

direction into the fluid tube. Do

not over-tighten.

There will be a small gap

between the flange of the fluid tip and the fluid
tube (see Figure 31). Insure the O-ring is
properly positioned when complete. Tighten to
25-30 lbs.- in. (2.83-3.4 Nm) torque.

> When removing fluid tip while turbine is still
installed, make sure to clean paint or fluid that
may have leaked or run onto the shaft or
threads.

Fluid Tube Removal/Replacement

Removal

Using the fluid tip/tube removal tool (A11229-
00), place the pinned end of the tool towards
the fluid tube retaining nut and engage the
pins into the holes. Turn the tool counter-
clockwise to remove (see Figure 32).

Replacement
Lubricate all O-rings with a suitable O-ring lub-
ricant. Push the fluid tube into the pocket of
the atomizer body. Seat the tube by pushing
while rocking the tube from side to side. In-
stall the fluid tube retaining nut over the tube.
Tighten the retaining nut firmly tight using the
removal tool in a clockwise direction. Tighten
to 65-75 lbs.- in. (7.34-8.47 Nm) torque.

Advertising