External trig setting, External trig setting” on – Teledyne LeCroy SierraFC M164 User Manual User Manual
Page 255

SierraFC M164 Protocol Analyzer User Manual
255
Software Settings
Teledyne LeCroy
External Trig Setting
The External Trig Setting dialog displays the External Trig Out Setting and External Trig In
Setting as High Active, Low Active, or Toggle.
Figure 3.79: External Trigger Setting Dialog
To display the External Trig Setting dialog, select Configuration > External Trig Setting.
External Trig Out Setting
The Analyzer can send a Low or High external signal anytime a trigger occurs. Select the
External Trig Out Setting: High Active, Low Active, or Toggle from High to Low or Low to
High once (3.3 V output).
Enter the External TrigOut pulse width.
External Trig In Setting
An external Low or High input signal can cause triggering. Select the
External Trig In Setting: High Active, Low Active, or Toggle from High to Low or Low to
High once (3.3 V output).
Note:
The nominal External Trigger voltage is 0.818 volts. Trigger In can work with 1 volt to 5 volts
input voltage.