Installation and self test, Inspecting the unit, Preparing the aed plus for use – ZOLL AED Plus Rev V User Manual

Page 25

Advertising
background image

ZOLL AED PLUS Administrator’s Guide

15

Installation and Self Test

This section describes the following functions to prepare the AED Plus for use:

• Inspecting the Unit

• Preparing the AED Plus for Use

• Using the Self Test Feature

• Installing or Replacing Batteries

• Identifying Battery Condition

Inspecting the Unit

Once unpacked, inspect the device for any signs of damage due to shipping. Check for accessories and

any other parts ordered.

Preparing the AED Plus for Use

To ensure that the AED Plus is functioning properly and is ready for use in an emergency situation,

the following set-up and checkout procedures should be performed before placing the device into

service and after each clinical use.

1.

Inspect all external surfaces of the unit to ensure that they are clean and free of structural damage

such as cracks, broken or missing parts.

2.

Inspect the electrode connector to ensure that there are no broken or missing connector pins.

3.

Verify that new CPR-D-padz, stat-padz® II or pedi-padz II to be used with the AED Plus are well

within their expiration date.

4.

Follow instructions provided with the new electrodes to pre-connect them to the electrode

connector on the device and pack them within the AED Plus cover.

NOTE If the electrodes are not connected to the AED Plus unit, the device will fail the self test and display

a red “X” in the status indicator window.

5.

If the status indicator displays a red X, install new batteries. (See “Installing or Replacing

Batteries” on page 18.)

6.

Close the top cover of the AED Plus unit and initiate a self test by pressing the Power Button.

Verify that the unit issues the UNIT OK voice prompt. This prompt indicates that the new batteries

and electrodes are properly installed and that the unit is ready for service.

7.

Verify that the AED Plus unit issues the appropriate “Adult Pads” or “Pediatric Pads” voice

prompt.

8.

Turn the AED Plus unit off.

9.

Wait 2 minutes. Verify that the green check symbol (

)

appears in the status indicator window

and that the unit does not emit a beeping tone.

10. Place the AED Plus unit in service.

11. Check the AED Plus unit periodically to ensure that the green check symbol (

)

appears in the

status indicator window.

NOTE If the status indicator displays a red X following completion of the above test, the AED Plus is not

ready for use and may be defective. Remove the AED Plus from service and consult the
Troubleshooting section on page 23 of this guide to help determine the problem.

Advertising