Guidance and Manufacturer’s Declaration - Electromagnetic
Emissions

Table 8: EMC Specifications

The AED Plus is intended for use in the electromagnetic environment specified below. The
customer or the user of the AED Plus should assure that it is used in such an environment.

Emissions test

Compliance

Electromagnetic environment -

guidance

RF emissions CISPR11

Group 1

The AED Plus uses RF energy only

for its internal function. Therefore,

its RF emissions are very low and

are not likely to cause any

interference in nearby electronic

equipment.

RF emissions CISPR 11

Class B

Harmonic Emission

IEC 61000 3-2

Not applicable

Voltage Fluctuations/Flicker

Emission

IEC 61000 3-3

Not applicable

Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and

put into service according to EMC information provided in this document.

The AED Plus is intended for use in the electromagnetic environment specified below. The
customer or the user of the AED Plus should assure that it is used in such an environment.

Immunity test

IEC 60601 test

level

Compliance

level

Electromagnetic environment -

guidance

Electrostatic

discharge (ESD)

IEC 61000-4-2

± 6 kV contact

± 8 kV air

± 6 kV contact

± 8 kV air

Floors should be wood, concrete or

ceramic tile. If floors are covered with

synthetic material, the relative

humidity should be at least 30%.

Electrical fast

transient/burst

IEC 61000-4-4

± 2 kV for power

supply lines

± 1 kV for input/

output lines

Not applicable

± 1 kV I/O

Surge

IEC 61000-4-5

± 1 kV differential

mode

+/- 2 kV common

mode

Not applicable

Not applicable