ZOLL AED PLUS Administrator’s Guide

Troubleshooting

The following table summarizes common error indications on the AED Plus, and their associated

corrective action. Return the AED Plus to ZOLL’s Technical Service Department if the AED Plus is not

working properly.

Table 5: Troubleshooting

Technical Problem

Recommended Action

Self test failed.

Initiate manual self test by pressing and holding the

ON/OFF Button for more than 5 seconds. Attempt to

repair the device by replacing the batteries or

electrodes. If the AED Plus fails test again, remove the

AED Plus from service and contact ZOLL Technical

Service.

CHANGE BATTERIES prompt.

Replace all batteries with new batteries at the same

time. Press the Battery Reset Button when prompted.

Red “X” in status indicator window.

Initiate manual self test by pressing and holding the

ON/OFF Button for more than 5 seconds.

Check to see if cable is attached properly to the AED

Plus or replace the electrodes.

Cycle power on the AED Plus by turning the unit off,

then on again.

Replace all batteries at the same time with new

batteries that are less than 1 year old. Press the Battery

Reset Button when prompted.

If the AED Plus still does not operate correctly, remove

unit from service and contact ZOLL Technical Service.

Beeping noise when AED Plus is off.

Remove AED Plus from service and replace batteries.

Replace all batteries at the same time with new

batteries. Press the Battery Reset Button when

prompted.

If beeping continues, contact ZOLL Technical Service.

PLUG IN CABLE prompt.

Check cable connection between electrodes and the

AED Plus.

ANALYSIS HALTED. KEEP VICTIM

STILL prompt.

Excessive artifact detected during ECG analysis.

Victim must be motionless during ECG analysis. Do

not touch the victim during analysis. Keep the victim

still. If the rescuer is using the AED Plus in an

emergency vehicle, bring the vehicle to a halt before

performing ECG analysis.

RELEASE SHOCK BUTTON prompt.

Release Shock Button, then press and hold Shock

Button until discharge occurs. If voice prompt

continues, contact ZOLL Technical Service.