ZOLL AED PLUS Administrator’s Guide

DEFIBRILLATOR (cont’d)

Maximum time from power on to
unit charged and ready to shock at
200J

22.6 seconds

Electrodes

ZOLL stat-padz II, CPR-D-padz or pedi-padz II

Built in Defibrillator Self Test

Included

CPR

*Metronome Rate: Variable 60 to 100 CPM

Depth: ¾" to 3"; 1.9 to 7.6 cm

Defibrillation Advisory

Evaluates electrode connection and patient ECG to determine if

defibrillation is required.

Shockable Rhythms: Ventricular fibrillation with average

amplitude>100 microvolts and wide complex ventricular

tachycardia with rates greater than 150 BPM (adult mode) and

200 BPM (pediatric mode). Refer to ECG Analysis Algorithm

Accuracy Section for sensitivity and specificity performance.

Electrode Patient Impedance
Measurement Range

0 to 300 ohms

Defibrillator Electrode ECG
Circuitry

Protected

ECG Bandwidth

2-30Hz

Display Format

Optional LCD with Moving Bar

Size: 2.6" x 1.3"; 6.6 cm x 3.3 cm

Viewing Time: 2.6 seconds

Display Sweep Speed

25 mm/sec

Battery Capacity

Typical new battery (20° C):

stand-by life with batteries installed: 5 years, or

300 ±5 continuous shocks: 250 ±5 shocks before “change

battery” indicator and 50 ±5 shocks after “change battery”

indicator; or

at least 13 hours of continuous Monitoring time

*

Testing reports validating performance and accuracy of CPR depth measurement capability,

adaptive metronome feature function and rescuer performance, and the PASS (Passive Airway

Support System) cover function are on file with ZOLL Medical Corporation and are available for

review. Contact ZOLL Technical Support to request a copy of the following report(s) if desired:

• Using the AED Plus Cover to Aid in Airway Patency

• Depth and Compression Rate Response of Real CPR Help

• AED Plus Real CPR Help Test Results.