ZOLL AED PLUS Administrator’s Guide

35

Recommended separation distances between portable and mobile RF communications
equipment and the AED Plus

The AED Plus is intended for use in an environment in which radiated RF disturbances are controlled. The
customer or the user of the AED Plus can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the AED Plus as
recommended below, according to the maximum output power of the communications equipment.

Separation distance according to frequency of transmitter

m

Rated maximum

output power of

transmitter

W

150 kHz to 80

MHz outside

ISM bands

150 kHz to 80

MHz in ISM

bands

80 MHz to 800

MHz

800MHz to 2.5

GHz

0.01

0.17

0.12

0.12

0.23

0.1

0.37

0.38

0.38

0.73

1

1.17

1.20

1.20

2.3

10

3.69

3.79

3.79

7.27

100

11.70

12.00

12.00

23.00

For transmitters rated at a maximum output power not listed above, the recommended separation

distance d in metres (m) can be determined using the equation applicable to the frequency of the

transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to

the transmitter manufacturer.

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2 The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6.765

MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to

40.70 MHz.

NOTE 3 An additional factor of 10/3 is used in calculating the recommended separation distance for

transmitters in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range

80 MHz to 2.5 GHz to decrease the likelihood that mobile/portable communications equipment

could cause interference if it is inadvertently brought into patient areas.

NOTE 4 These guidelines may not apply in all situations. Electromagnetic propagation is affected by

absorption and reflection from structures, objects and people.

d =

P

3.5

3

[

]

d =

P

12
10

[

]

d =

P

12
10

[

]

d =

P

23

10

[

]

Advertising