To remove the battery cover, insert a small tool (e.g., flat blade screw driver) into the two slots on

the rear of the unit to depress the latches, then insert the tool into the groove on the bottom to lift

the cover (see Figure 7).

Figure 7: Removing the Battery Compartment Door

2.

Remove all batteries at once and dispose of batteries properly. Place new batteries into the battery

bank, observing battery polarity markings and making sure that all batteries are securely seated and

properly oriented. After placing the first 5 to 9 batteries in the battery well, the

INSTALL BATTERIES audio prompt reminds you to install the remaining batteries in the battery

compartment.