ZOLL AED PLUS Administrator’s Guide

37

Figures 9 through 14 show the Rectilinear Biphasic waveforms that the AED Plus defibrillator

produces when it discharges into loads of 25, 50, 75, 100, 125, 150, and 175 ohms at each energy

setting (200, 150, 120, 85, 70, and 50 joules).

The vertical axis shows the current in amperes (A); the horizontal axis shows the duration in

milliseconds (ms).

Figure 9: Rectilinear Biphasic Waveforms at 200 joules

Figure 10: Rectilinear Biphasic Waveforms at 150 joules