ZOLL AED PLUS Administrator’s Guide
37
Figures 9 through 14 show the Rectilinear Biphasic waveforms that the AED Plus defibrillator
produces when it discharges into loads of 25, 50, 75, 100, 125, 150, and 175 ohms at each energy
setting (200, 150, 120, 85, 70, and 50 joules).
The vertical axis shows the current in amperes (A); the horizontal axis shows the duration in
milliseconds (ms).
Figure 9: Rectilinear Biphasic Waveforms at 200 joules
Figure 10: Rectilinear Biphasic Waveforms at 150 joules