Identifying battery condition – ZOLL AED Plus Rev V User Manual

Page 31

Advertising
background image

ZOLL AED PLUS Administrator’s Guide

21

3.

After installing new batteries, press the Battery Reset Button inside the battery compartment when

prompted (see Figure 8). Pressing the button resets the battery usage indicator to full charge.

NOTE If you do not press the Battery Reset Button in the battery well within 15 seconds after installing all

batteries, the AED Plus assumes that the batteries installed in the device were temporarily removed,
and are not fully charged.

NOTE Because Lithium Manganese Dioxide battery cells do not contain toxic materials, they may be disposed

of in general trash after discharge or when properly protected against shorting between terminals.

Identifying Battery Condition

Battery capacity depletes during standby operation of the unit, while the unit is operating, and as a

result of each defibrillation. Battery capacity also gradually diminishes over a shelf life of years

without use. The AED Plus monitors energy remaining in the installed batteries. When battery

capacity is low or depleted, the AED Plus will not function to specification. When a low battery

condition occurs, the AED Plus:

• emits an audible alarm or “beep” once every minute, if the AED Plus is off

• issues the CHANGE BATTERIES audio prompt, if the AED Plus is on

• displays a red “X” in the status indicator window, indicating that the batteries are at low capacity

or that the AED Plus has failed other self tests.

CAUTION! You MUST replace all 10 batteries at once. Do not replace individual batteries. The AED

Plus cannot detect whether all batteries or only a few batteries have been replaced. Do
not place used batteries into the AED Plus. Using less than fully charged batteries
can affect the unit when performing a rescue
. DO NOT press the Battery Reset
Button if all the batteries are not new. The AED Plus then assumes that they are the same
batteries that were just removed.

Table 3: Battery Condition

Battery Condition

Indications

Correction

Low Battery with AED

Plus off.

Audible beep from AED Plus

once every minute.

Replace batteries.

Low Battery during power

up self test.

CHANGE BATTERIES prompt

(when AED Plus is powered on)

Replace batteries.

Low Battery or other self

test failure with AED Plus

powered off or during self

test.

Status Indicator has red “X”

indicating failure to operate

(when off).

Replace batteries. Check or

replace electrodes. If red “X”

remains, return to ZOLL

Technical Support for service.

Low Battery with AED

Plus powered on.

CHANGE BATTERIES prompt

(unit powered on).

Replace batteries as soon as

possible.

Dead Battery

Status Indicator has red “X”

indicating failure to operate

(when off).

Replace batteries. If red “X”

remains, return to ZOLL

Technical Support for service.