Figure 3-1 autopulse system nimh battery, 1 expected autopulse nimh battery life, 2 handling new autopulse nimh batteries – ZOLL AutoPulse Rev A User Manual

Page 16: Figure 3-1 autopulse system nimh battery -2

Advertising
background image

Page 3-2

P/N. 12457-001 Rev. 2

Power System User Guide

Figure 3-1 AutoPulse System NiMH Battery

3.1

Expected AutoPulse NiMH Battery Life

The expected service life of the AutoPulse NiMH Battery is 100 full charge/discharge cycles or 2 to 4
years depending on battery maintenance and usage patterns. ZOLL recommends that customers plan to
purchase batteries on intervals that reduce the likelihood that they will have all batteries due for
replacement simultaneously. A specific replacement interval may be difficult to establish until use and
charging patterns are well established. Some some customers may want to replace batteries on a
preventative basis regardless of the Battery’s capacity or remaining life.

Note:

The NiMH Battery will not operate after 100 full charge/discharge cycles.

3.2

Handling New AutoPulse NiMH Batteries

New AutoPulse NiMH Batteries should be unpacked immediately and not stored for any extended
period. Before putting new AutoPulse NiMH Batteries into service they must be placed into the
AutoPulse Multi-Chemistry Battery Charger, and run through a Test-Cycle. A Test-Cycle is initiated by
the Charger automatically, and no user action is required. Once a test cycle is initiated, do not remove
the AutoPulse NiMH Battery from the Charger until the Test-Cycle is complete. This process could take
up to 12 hours.

Red LED

Amber LED

Green LED

Status Check Button

Power and Communications

Remove Protective
Plastic Cap before use