Page 4-4

P/N. 12457-001 Rev. 2

Power System User Guide

4.

AutoPulse Li-Ion batteries placed in a charging bay are automatically charged and tested for
minimal performance (Charging LED illuminated) in less than 4

¼

hours.

5.

AutoPulse NiMH batteries placed in a charging bay are automatically charged and tested for
minimal performance (Charging LED illuminated) in less than 6

¼

hours.

Note:

Do not remove a Battery from the Multi-Chemistry Battery Charger until its charging
completes, or the Battery’s run time will be reduced.

6.

The AutoPulse Multi-Chemistry Battery Charger automatically conducts a Performance Test on
both the AutoPulse Li-Ion Battery and the AutoPulse NIMH Battery to ensure each Battery meets
performance standards. The Performance Test occurs every time a Battery is placed into the
AutoPulse Multi-Chemistry Battery Charger.

7.

Once the green LED on the Charger illuminates

the Battery is fully charged and has successfully passed

the Performance Test. The Battery is ready to use.

Table 4-1 AutoPulse Multi-Chemistry Battery Charger Status LEDs (Page 1 of 2)

Multi-

Chemistry

Battery

Charger

Mode

Multi-Chemistry

Battery Charger

Status LEDs Used

Definition

Action

Charge

Charging (yellow LED)

The Battery is charging.

Leave the Battery in the Charger until
the Ready (green) LED illuminates.

Test-Cycle

Charging (yellow LED)

Test (amber LED)

Typical Test-Cycles last up to 12
hours.

Leave the Battery in the Charger until
the Test-Cycle completes and the
ready (green) LED illuminates.

Ready

Ready (green LED)

The Battery is fully charged and
has successfully passed the
Performance Test. The Battery
is ready to use.

Leave the Battery in the Charger
to ensure that the Battery is fully
charged when needed or

Install in the AutoPulse Platform
or

Store in a cool location.

Fail

Fail (red LED)

The Charger was unable to
charge the Battery or

the Battery has failed the
Performance Test or

the Battery has failed a test
cycle or

the Battery has reached its
end of life.

Remove and re-insert the Battery into
the Charger. If the Fail LED remains
illuminated, contact ZOLL.

Advertising