Appendix c technical specifications, C.1 li-ion battery physical and environmental, Category – ZOLL AutoPulse Rev A User Manual

Page 43: Specifications

Advertising
background image

P/N. 12457-001 Rev. 2

Page C-1

Power System User Guide

Appendix C Technical Specifications

The specifications provided in this appendix apply to the AutoPulse Power System.

C.1

Li-Ion Battery Physical and Environmental

Table C-1 Li-Ion Battery Specifications (Page 1 of 2)

Category

Specifications

Manufacturer

ZOLL Circulation

Model Number

8700-0752-01

Size (L

WH)

11.5 in. by 3.2 in. by 2.2 in. (29.2 cm by 8.1 cm by 5.7 cm).

Weight

3.0 lbs. (1.3 kg).

Type

Rechargeable Lithium-Ion (LiFePO

4

)

Battery voltage (nominal)

36.3V

Capacity

2300 mAh (typical)

Current (maximum)

30 A continuous, 48 A pulse (96 ms max)

Initial Battery run time
(nominal patient)

30 minutes (typical)

Maximum Battery charge
time

Less than 4¼ hours at 77ºF (25ºC)

Battery Test-Cycle time

Less than 12 hours per Test-Cycle session

Recommended replacement
interval

3 years from date of manufacture

Note:

The Battery will not operate after 5 years from date of manufacture.

Operating temperature

+32º to +113ºF (0º to +45ºC) ambient installed in device

Charge temperature

+41º to +95ºF (5º to +35ºC) ambient (68º to 77ºF [20º to 25ºC] preferred)

Storage temperature

-4º to +113ºF (-20º to +45ºC) ambient for up to six months with charging every
four weeks, starting with a fully charged Battery.

Operating altitude

0 to 15,000 ft. (0 to 4,572 m)

Enclosure protection

Meets IP24 per IEC 60529

Shock

Meets IEC 60068-2-27 Basic Environmental Testing Procedures – Shock (50g,
11ms pulse, half sine wave)

Vibration

Meets IEC 60068-2-6 Basic Environmental Testing Procedures (10 to 150 Hz,
10 m/s

2

)

Meets IEC 60068-2-64 Basic Environmental Testing Procedures – Random
Vibration Broad Band – General Requirements (f1:20, f2:2000, ASD 0.05)

Free fall

Meets IEC 60068-2-31 Basic Environmental Testing Procedures – Free Fall –
Procedure 1.

Electrostatic discharge

Meets IEC 61000-4-2, Level 3

Advertising