Battery Rotation

After every use, at the beginning of a shift, or at least once every 24 hours, the Battery in the AutoPulse
should be replaced with a fully charged Battery.

Figure 5-5 Four Battery rotation

Four Battery Rotation—to be done after every use and/or once per shift (see Figure 5-5 above).

•

Remove the Battery from the AutoPulse, and place it into the Charger.

•

Remove the spare Battery, and place it into the Charger.

•

Take two fully charged Batteries from the Charger, check for green LEDs on each Battery, and then
place one into the AutoPulse, and use the second as a spare.

•

Power on the AutoPulse and ensure no faults are displayed.