Table 4-1 battery charger status leds -3, Table 4-1 sho – ZOLL AutoPulse Rev C User Manual

Page 21

Advertising
background image

P/N. 10762-001 Rev. 8

Page 4-3

Battery Charger User Guide

5.

When the READY LED illuminates, either:

Remove the charged Battery from the Battery Charger by gripping the Battery while pressing
up on the finger latch to disengage the locking bar. Pull the Battery straight out until it fully
clears the charging bay.

Leave the Battery in the Battery Charger to maintain peak capacity.

Note:

Do not remove a Battery from the Battery Charger until its charging is complete or the
Battery’s runtime will be reduced.

Note:

Recently-charged Batteries can be warm to the touch. This is a consequence of normal
operation.

Table 4-1

Battery Charger Status LEDs

Battery
Charger
Mode

Battery Charger
Status LEDs Used

Definition

Action

Charge

CHARGING (yellow
Battery Charger LED)

The Battery is charging.

Leave the Battery in the Battery
Charger until the READY (green)
LED illuminates (less than 4π hours).

Test-Cycle

CHARGING (yellow
Battery Charger LED)

TEST (amber Battery
Charger LED)

Typical test-cycles last
approximately 10 hours. If
Batteries require additional
testing, the test-cycle may take
approximately 22 hours.

Leave the Battery in the Battery
Charger until the test-cycle
completes. At completion, the Battery
Charger indicates the Battery’s status:
READY or FAIL.

Ready

READY (green Battery
Charger LED)

The Battery is charged and
ready to use.

Leave the Battery in the Battery
Charger to ensure that the
Battery is fully charged when
needed.

Remove and store in the
AutoPulse Platform.

Remove and store in a cool
location.

Fail

FAIL (red Battery
Charger LED)

The Battery Charger is currently
unable to charge the Battery.

Refer to Table A-2 for more
information.

Idle

The Battery Charger is unable to
recognize the Battery.

Remove and re-insert the Battery. If
the status is still Idle, refer to
Appendix A, “Troubleshooting” for
more information.

Advertising