Appendix a, “troubleshooting” for more information, Appendix a troubleshooting – ZOLL AutoPulse Rev C User Manual

Page 27

Advertising
background image

P/N. 10762-001 Rev. 8

Page A-1

Battery Charger User Guide

Appendix A Troubleshooting

This Appendix details symptoms, possible causes, and recommended actions for difficulties you might
have with your AutoPulse Power System. Table A-1 provides troubleshooting procedures for the Battery.
Table A-2 provides troubleshooting procedures for the Battery Charger.

Table A-1

Battery Troubleshooting Procedures

Symptom

Possible Cause

Recommended Action

Battery Status Check
LEDs do not illuminate.

The Battery’s status is
unknown.

Place the Battery in one of the Battery Charger’s
charging bays.

1.

If the Battery Charger’s CHARGING LED is
illuminated, the Battery Charger is attempting to
restore the Battery. Refer to Chapter 4,
“Operating the Battery Charger” for more
information.

2.

If the Battery Charger’s FAIL LED is
illuminated, the Battery has failed. Replace the
Battery. Refer to Section 5.2.2, “Reaching the
End of Battery Service Life” for more
information.

Battery will not fully insert
into the Battery Charger.

Protective plastic cap
was left on the Battery.

The Battery may be
damaged.

The Battery Charger’s
charging bay might be
obstructed.

Remove protective cap.

Inspect the guide rails around the connector for
damage. If the guides are damaged, replace the
Battery.

Inspect the Battery connector for damage. If the
connector is damaged, replace the Battery.

Unplug the alternating current (AC) from the
Battery Charger. Check the Battery Charger’s
charging bay to ensure that no debris has
accumulated in the bay.

The Battery will not fully
insert into the AutoPulse
Platform.

Protective plastic cap
was left on the Battery.

The Battery may be
damaged.

The AutoPulse
Platform’s Battery
compartment might be
obstructed.

Remove protective cap.

Inspect the guide rails around the connector for
damage. If the guides are damaged, replace the
Battery.

Inspect the Battery connector for damage. If the
connector is damaged, replace the Battery.

Check the AutoPulse Platform’s battery
compartment to ensure that no debris has
accumulated in the bay.